13 The Configuration
52
13.2 Configuration - Weighing parameters
Refresh rate:
fast,
normal
, slow, extra slow
The value set defines the period after which each new measurement is displayed.
For the definition of this period, the quality of the moisture analyser location is crucial. The stability control
must also be set appropriately, see next point „Environment“.
Environment:
stable,
normal
, unstable
Adjust the weighing system to the quality of the dryer location, see also chapter 5.4 "Choosing a Suitable
Location".
Autostandby time: Off
, 30 Seconds, 1 Minute, 5 Minutes, 30 Minutes
Defines the period of non-use before the dryer automatically switches into the energy-saving mode.
The Autostandby only works with the automatic zero-correction enabled („Autozero“).
To re-start the dryer, press any key, put a weight on the dryer or execute a remote control command via the
interface.
Autozero: On
, Off
Automatic zero-correction.
The dryer holds a stable zero (e.g. even with fluctuations in room temperature).
Quicktare:
On,
Off
The dryer executes a fast tare, even at not yet stabilized weight.
13.3 Configuration - Adjustment
For the adjustment of the dryer, see chapters 5.8 "Weight Adjustment" and 16.2 "Adjustment". The setting
depends on the dryer model.
Adjustment
Mode:
Off, External, Ext. user-def.weight,
Internal
, Automatic
- Adjustment closed
- External adjustment
- External adjustment with a user-defined calibration weight, see „[
Def.-Weight]
“
- Adjustment with the internal calibration weight
- Automatic adjustment on time, temperature or time and temperature (adjustment is disabled if moisture an-
alyser is heating, choose a time when usually you are not doing analysis)
[User-def.weight:]
0.000 g, n.nnn g.
(only available in Adjustment Mode External Def.-Weight)
Defines an external calibration weight
Configuration - Weighing parameters
Refresh rate
normal
Environment
normal
Autostandby time
Off
Autozero
On
Quicktare
Off
ESC
"
!
E
Configuration - Adjustment
Adjustment mode
#
Internal
[Calibration weight]
0.000 g
[Weight ID]
____
[Automatic mode]
Time & Temperature
[Temperature diff.]
2 C
[Time]
06:00:00
Protocol
On
ESC
"
!
E