13 The Configuration
■
53
En
gl
is
h
[Weight ID:] ____
(only available in Adjustment Mode External Def.-Weight)
Sets an alphanumeric identification to the external calibration weight
[Automatic Mode:] Time & Temperature
, Temperature, Time
(only available in Adjustment Mode Automatic)
- Automatic adjustment on time and temperature
- Automatic adjustment on temperature
- Automatic adjustment on time
[Temperature:] 2 C
(only available in Adjustment Mode Automatic)
Defines how much the temperature has to change to start the automatic adjustment. Pre-set is a temperature
change of 2. The unit is about degree Celsius.
[Time:]
06:00:00
(only available in Adjustment Mode Automatic)
Defines the time of day on which to start the automatic adjustment
Protocol: On
, Off
Executes, after a adjustment, a printout of the adjustment protocol
13.4 Configuration - Protocol
Just pressing «
» transmits the current measuring result via interface to the peripherals. Setting up the
protocol, you can
• define under what conditions this has to happen
• define the format of the value (measuring result) in the printout
• choose various additional information to be printed out.
The main screen to setup the protocol:
Autostart:
Define
Defines the automatic start of the printout after the moisture analyser has been switched on. An ‘*‘ is dis-
played if the autostart is enabled
Mode:
Define
Defines the mode of the printout, e.g. stable or after a loadchange, and so on
Value format:
Define
Defines the format of the weight value in the printout
Header:
Define
Defines the format of a header in the printout. An‘*‘ is displayed if the Header is enabled
Footer:
Define
Defines the format of a footer in the printout. An‘*‘ is displayed if the Footer is activated
NOTE
The protocol set in configuration regards only the weighing values.
The protocol of analysis work with the protocol setting in the context menu (see 12.3 "Protocol")
Configuration - Protocol
Autostart
#
Define
Mode
Define
Value format
Define
Header
Define
Footer
Define
ESC
"
!
E