8 Create a new Method
■
21
En
gl
is
h
If stop time and stop mode are OFF you will be allerted: No stop mode defined
Unit:
Set the calcuation and unit of the result.
Explanation of the variables used:
– DW: Dry weight (weight value at the end of the interval or measurment)
– MW: Moist weight (weight value at the start of the measurement)
•
Unit:
Choose the calcuation and unit.
•
Result Base:
Select the value used as moist weight MW (
is only avaialable in Interval 2 and Interval 3
)
•
Result correction factor:
Off
, On
Choose to correct the result and enter the correction factor.
Caluations for Loss % and Loss g/kg using Sutrahend and Divisor which is only available in Interval 2 and
Interval 3:
Explanation of the variables used
– DW: Dry weight (weight value at the end of the interval or measurment)
– SW: Weight used as subtrahend in calculation
QW: Weight used as divisor in caculation
•
Subtrahend
: Select the value to use as subtrahend SW
•
Divisor
:Select the value to use as divisor QW
•
Result correction factor
:
Off
, On
Choose to correct the result and enter the correction factor
Unit
Calculation
Loss % (moisture):
Resiual % (dry mass):
ATRO moisture:
ATRO dry mass:
Loss g / kg [‰]:
Residual g / kg [‰]:
Weight loss in g:
Residual in g:
Loss % positive (moisture):
Unit
Calculation
Loss % (moisture):
Loss g / kg [‰]:
MW DW
–
MW
--------------------------- 100%
⋅
–
DW
MW
----------- 100%
⋅
MW DW
–
DW
--------------------------- 1
⋅
·
00%
–
MW
DW
----------- 100%
⋅
MW DW
–
MW
---------------------------
–
1000
⋅
DW
MW
----------- 1000
⋅
MW DW
–
DW
MW DW
–
MW
--------------------------- 100%
⋅
DW SW
–
QW
------------------------- 100%
⋅
DW SW
–
QW
------------------------- 1000
⋅