5 Setup
■
11
En
gl
is
h
5.4
Choosing a Suitable Location
The dryer location must be chosen in such a way as to guarantee perfect operation of your instrument, so
that the allowable ambient conditions and prerequisites are met and maintained:
• Put the moisture analyser on a solid, firm and preferably vibration-proof, horizontal base
• Make sure that the dryer cannot be shaken or knocked over
• Protect from direct solar radiation
• Avoid drafts and excessive temperature fluctuations
•
Leave enough clear space around the instrument to prevent a build-up of heat.
5.5
Connecting the Moisture Analyser to the Mains
The following safety recommendations must be observed when connecting the instrument to the mains:
5.6
Safety measures
The moisture analyser may only be plugged into a grounded wall outlet. The safety effect may not be undone
by using an extension lead without a ground wire. If the voltage is from a power source without a protective
earth terminal, arrange for an electrician to create a comparable level of protection in accordance with local
installation regulations.
NOTE
With difficult conditions (where the dryer may be easily shaken or subject to vibration) the dryer can nev-
ertheless provide accurate results through suitable adjustment of the stability control (see chapter 13.2
"Configuration - Weighing parameters").
DANGER
The instrument may only be operated using the original mains cord supplied.
If the mains cord supplied is not long enough, only use an extension cord fitted with a protective earth
conductor.
Plug the mains cord into a socket which has been installed in accordance with regulations and is fitted
with a PE terminal..
For technical reasons, the heating unit is designed in the factory to accommodate a voltage of 230 V or
115 V and in accordance with your order. Check that the setting matches the local setting?
If not the
instrument
must be connected to the mains supply. Contact the Customer Service.