16 Maintenance and Servicing
68
16 Maintenance and Servicing
16.1 Utility program
• Hold the Tara-Key to open the utility program
Balance Adjustment
Balance adjustment as definied in the configuration menu.
Balance Adj. History
You get the information on the last adjustments.
Temperature Calibration
Two point temperature calibration or adjustment.
Temp. Adj. History
You get the information on the last temperature adjustments.
Auto Repeatability Test
Repeatabilty test with internal calibration weight.
Test measurement
Helps you to develop a method with the help of reference material.
16.2 Adjustment
The adjustment of the balance is fixed in the Configuration menu ( 5.8 "Weight Adjustment" and 13.3 "Con-
figuration - Adjustment").
Possible types of adjustment:
• External adjustment by means of ICM (Intelligent Calibration Mode)
• External adjustment with freely selectable weight
• Internal adjustment
• Automatic adjustment
To manually perform a adjustment proceed as follows:
• Hold «
0/T
» until the utility program is shown.
• Select „Balance Adjustment“ and press
{
E
}
The adjustment starts.
16.2.1 External adjustment
Adjustment weights in steps of 10 g can be used, where the calibration weight must correspond to the pre-
cision of the dryer.
For external adjustment with user-definable weight, „Adjustment Mode = External“ must be selected in the
Configuration menu (see chapter 13.3 "Configuration - Adjustment").
Utility program
#
Balance Adjustment
Balance Adj. History
Temperature Calibration
Temp. Adj. History
Auto Repeatability Test
Test measurement
ESC
"
!
E
NOTE
The adjustment can be interrupted at any time by pressing «
ON/OFF
».