User manual
PONSEL
Ref.: NOTICE_ACTEON_2020_v003
V : 003
50/57
After having reset the factory default parameters, ACTEON 2020 will be configured with the following
parameters:
Sensor response averaging
10 measurements
Time base for trend line display
1 minute
4-20mA output adjustment
Redox output: The 4mA stop is set to -500mV and the
20mA stop is set to 500mV
Temp. °C output: The 4mA stop is set to -5°C and the
20mA stop is set to 35°C.
Relay output adjustment
The relays are configured in mode 1:
- R1: threshold: 0mV hysteresis: 5mV above the threshold,
the R1 relay is active
- R2: threshold: 50.00°C; hysteresis: 0.10 above the
threshold, the R2 relay is active
Temperature measurement unit
°C
Language
English
10 Information menu
The INFO window provides information on the type and version of the device. The software and hardware
versions will be requested when you call PONSEL Technical support or After-sales service.
INFO
ACTEON 2020 ORP
Temperature
Software version : XX.XX
Hardware version : XX.XX
Esc: Escape menu