Adjust the 4mA outputs and press ENTER.
Use the
and
keys to increase
and decrease the 20mA value.
Use the
and
keys to select the
ORP 4-20mA output or the
temperature 4-20mA output.
CONFIG
4-20mA OUTPUTS
CURRENT GENERATION OF 20mA
Enter the value
read on the ORP
4-20mA output
20.00
mA
Enter the value read
on the temperature
4-20mA output
20.00
mA
Adjust the 20mA outputs and press ENTER.
CONFIG
WHAT DO YOU WANT TO DO?
Averaging sensor response
Graphic screen
4-20mA outputs adjustment
Relay adjustment
Measure units
Language
Factory default value reminder