9.3.2 Calibrating 4-20mA outputs
Calibrating the 4-20mA outputs improves their level of accuracy by taking into account the device load.
Use the
and
keys to select
the configuration required.
Select the
4-20mA output calibration
menu and
press ENTER.
CONFIG
4-20mA OUTPUTS
4-20mA output adjustment
4-20mA output calibration
Use the
and
keys to increase
and decrease the4mA value.
Use the
and
keys to select the
ORP 4-20mA output or the
temperature 4-20mA output.
CONFIG
4-20mA OUTPUTS
CURRENT GENERATION OF 4mA
Enter the value
read on the ORP
4-20mA output
4.00
mA
Enter the value read
on the temperature
4-20mA output
4.00
mA
ESC: Escape menu
ENTER: To continue
Connect an ammeter to the 4-20mA outputs and press
ENTER.
CONFIG
Procedure:
Insert an ammeter on 4-20mA
outputs.
ENTER: To continue
ESC: Escape menu
AMMETER
LOAD