User manual
PONSEL
Ref.: NOTICE_ACTEON_2020_v003
V : 003
27/57
Use the
and
keys to adjust the
water temperature value.
Press the ENTER key to confirm the first
calibration point.
CALIBRATIO
N
TEMPERATURE SENSOR
Immerse the sensor in
water and ice
Standard:
Measure:
°C
°C
ESC: Cancel the procedure
ENTER: Validate the zero
0.00
0.15
WAIT: Nonstable measurement
AWAIT MEASUREMENT STABILISATION
Use the
and
keys to adjust the
water temperature value.
Press the ENTER key to confirm the second
calibration point.
CALIBRATION
TEMPERATURE SENSOR
Immerse the sensor in
water at ambient temperature
Standard:
Measure:
°C
°C
ESC: Leave the current slope
ENTER: Validate the new slope
20.00
20.20
WAIT: Nonstable measurement
AWAIT MEASUREMENT STABILISATION
Immerse the sensor in water at ambient temperature.
If the first calibration point is incorrect, an error
message window appears (see "Error message
information" chapter §6.3.1)
No
Yes
Calibration
Correct