6.1.5 Temperature sensor slope adjustment
If you observe a small measurement error, you may only adjust the slope of your sensor by carrying out the
following procedure:
CALIBRATION
SENSORS CALIBRATION
OPR SENSOR
Gain adjust
Complete calibration
Theoretical calibration
TEMPERATURE SENSOR
Gain adjust
Complete calibration
Theoretical calibration
ENTER: Validate the choice
Use the
and
keys to select
the type of calibration and the
probe to be calibrated.
Select the
Gain adjust
menu and press ENTER.
Use the
and
keys to adjust the
ambient water temperature value.
Press the ENTER key to confirm slope
adjustment.
CALIBRATION
TEMPERATURE SENSOR
Immerse the sensor in
water at ambient temperature
Standard:
Measure:
:
°C
°C
20.00
20.20
WAIT: Nonstable measurement
AWAIT MEASUREMENT STABILISATION
Immerse the sensor in water at ambient temperature.
ESC: Leave the current slope
ENTER: Validate the new slope