8 VIEWING THE SENSOR CALIBRATION REPORT
The calibration logger is used to check the reports of the most recent calibrations.
Follow the instructions below to access this information:
List of diagnostic messages from the last calibration :
Message
Explanation
CORRECT
ACTEON 2000 has been calibrated correctly.
THEORETICAL coef.
ACTEON 2000 is using the theoretical coefficients (slope and offset); this
message is displayed after a theoretical calibration.
THEORETICAL slope
ACTEON 2000 is using the theoretical slope.
THEORETICAL offset
ACTEON 2000 is using the theoretical offset.
INCORRECT slope
ACTEON 2000 is incorrectly calibrated because the slope calculated during
calibration is incorrect (in this case ACTEON 2000 is using the last correctly
calculated slope).
INCORRECT offset
ACTEON 2000 is incorrectly calibrated because the offset calculated during
calibration is incorrect (in this case ACTEON 2000 is using the last correctly
calculated offset).
Use the
and
keys to
navigate the ACTEON menu.
Select the menu and press ENTER.
LOG
CALIB LOG
CALIBRATION LOGGER
Calibration of
redox sensor
CORRECT
Standard 1: 0mV
Standard 2: 240mV
Offset: 1mV
Slope: 110%
Name: Ponsel
Calibration of
Temperature sensor
CORRECT
Standard 1: 0.01°C
Standard 2: 20.00°C
Offset: 1.65°C
Slope: 110%
Name: Ponsel
Esc: Escape menu
20.00
°C
-24 hrs
290
MEASURE
mV
INFO
CONF
HIST CAL
LOG
MEAS
Mode1 R1
R2
ACTEON 2020 ORP
mV
200
300