6.1.6 Returning to temperature measurement theoretical calibration
Theoretical calibration is carried out using the theoretical coefficients (PT100 theoretical slope and offset
zeroing).
Follow the instructions below to carry out theoretical calibration:
CALIBRATION
SENSORS CALIBRATION
ORP SENSOR
Gain adjust
Complete calibration
Theoretical calibration
TEMPERATURE SENSOR
Gain adjust
Complete calibration
Theoretical calibration
ENTER: Validate the choice
Use the
and
keys to select
the type of calibration and the
probe to be calibrated.
Select the
Theoretical calibration
menu and
press ENTER.
Use the
and
keys to select the
procedure confirmation.
Select
Yes
and press ENTER to return to the
theoretical coefficients.
CALIBRATION
THEORETICAL CALIBRATION
Esc: Cancel the procedure
Enter: Validate the choice
ATTENTION, you will erase your calibration
coefficients to return to the theoretical
coefficients!
Yes
No