6.2 Redox sensor calibration error message
6.2.1 Automatic zero measurement error
If you observe an error during automatic zero measurement: Contact the after-sales service (see last page)
6.2.2 240mV or 470mV buffer solution calibration error
If an error occurs during calibration with the 240mV or 470mV buffer :
1)
Check if the buffer is fouled or out-of-date.
2)
Check the connection between the sensor and ACTEON 2000.
3)
Clean the platinum electrode (the sensor's response will be weak if the platinum electrode is soiled).
6.3 Temperature sensor calibration error message information
6.3.1 0°C calibration error
CALIBRATION
Esc: Escape menu
ORP SENSOR
CALIBRTION IMPOSSIBLE
Slope:
NON CORRECT
CALIBRATION
TEMPERATURE SENSOR
CALIBRATION IMPOSSIBLE
Offset:
NON CORRECT
Esc: Escape menu
CALIBRATION
ORP SENSOR
CALIBRATION IMPOSSIBLE
Offset:
NON CORRECT
Esc: Escape menu