6.1.3 Returning to REDOX measurement theoretical calibration:
Theoretical calibration is carried out using the theoretical coefficients.
Follow the instructions below to carry out theoretical calibration:
20.00
°C
-24 hrs
290
MEASUR
E
mV
INFO
CONF
HIST CAL
LOG
MEAS
Mode1 R1
R2
ACTEON 2020 ORP
mV
200
300
CALIBRATION
SENSORS CALIBRATION
ORP SENSOR
Gain adjust
Complete calibration
Theoretical calibration
TEMPERATURE SENSOR
Gain adjust
Complete calibration
Theoretical calibration
ENTER: Validate the choice
Use the
and
keys to select
the type of calibration and the
probe to be calibrated.
Select the
Theoretical calibration
menu and
press ENTER.
THEOR CAL
!
Indicates that the device has
been calibrated with the
theoretical coefficients.
Use the
and
keys to select the
procedure confirmation.
CALIBRATION
THEORETICAL CALIBRATION
Esc: Cancel the procedure
Enter: Validate the choice
ATTENTION, you will erase your
calibration coefficients to return to
the theoretical coefficients!
Yes
No
Select
Yes
and press ENTER to return to the
theoretical coefficients.