3-23
3 First Steps
Objective
positioner (with
MSA-A-TMS
option)
4. Connect the protective earth cable to the objective positioner as shown in
5. Tighten the screw [D] with a torque of 1.1 Nm.
6. Make sure that the contact resistance is < 0.1
Ω
at 25 A at all connection
points relevant for mounting the protective earth cables.
Figure 3.17: Connection for the protective earth cable on the objective positioner
$
%
&
'
A
M4 contact washer
B
M4 cable lug
C
Washer Ø 4.3 mm for M4 screw
D
M4 x 8 head screw (special screw with a very flat head)