Planmeca Proline XC X-ray unit C-1
Chapter
Technical Manual
C
HELP & ERROR MESSAGES
1 HELP MESSAGES
The following is a list of the help messages. Touch the OK field to clear the help message
from the display.
2 ERROR MESSAGES
The purpose of the error messages is to make fault-finding easier. The list of the last 50 error
occurrences is of a great help. Please refer to the section 2.5 “Error history” on page B-12
how to display the error list.
HELP CODE
HELP MESSAGE EXPLANATION
HE 06
The temple support motor has been inactivated. The support movement
does not operate.
HE 15
The X-ray unit is not communicating with the PC, e.g. the Dimaxis soft-
ware is not running.
HE 16
The patient is not selected in the Dimaxis software.
HE 17
The sensor head is not in its position.
HE 18
In the cephalometric mode there is also a sensor head in the rotating
unit in addition to the sensor head attached to the cephalostat. Remove
the sensor head from the rotating unit.
HE 19
The unit exposure mode (panoramic/cephalometric) differs from the
mode selected in the Dimaxis software.
ERROR CODE
ERROR MESSAGE EXPLANATION
Er 00
The exposure switch was released too early during exposure.
Er 01
Short loss of power or drop in line voltage.
Er 07
The rotating unit is not in the ready position.
Summary of Contents for Proline XC
Page 1: ...technical manual 10010942_10 EN...
Page 2: ......
Page 8: ...TABLE OF CONTENTS 6 Planmeca Proline XC with Dimax3 Technical manual...
Page 253: ...Planmeca Proline XC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Page 254: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline XC X ray unit Technical Manual...
Page 256: ......
Page 257: ......