Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
E-28 Planmeca Proline XC X-ray unit
CALIBRATING PANORAMIC SENSOR HEAD
Technical Manual
The Dimax2Tool window appears.
Figure 52
Calibrating the sensor head
Select Proline from the Settings / Type pulldown menu.
Figure 53
The calibration must be performed with both 3x3 (enhanced resolution) and 4x4 (normal res-
olution) binning. Select the binning from the Setting / Binning pulldown menu.
Figure 54
Click the Dimax3 panoramic exposure button. A dark current exposure (i.e. an exposure with-
out radiation) must be taken first and a message window shown below appears.
Figure 55
Dimax3 panoramic
exposure