Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
E-52 Planmeca Proline XC X-ray unit
PATIENT POSITIONING LIGHTS
Technical Manual
5 PATIENT POSITIONING LIGHTS
5.1
Checking the lights
Place the Frankfort plane alignment tool into position on top of the ball phantom.
Figure 103
Touch either of the Layer arrow fields to switch the positioning lights on and then a second
time to drive the patient positioning mechanism to 0 if it is not already there.
Figure 104
START
END
PLANMECA
No.50971
X-LINE
Y-
LIN
E
PL
AN
ME
CA
O
Y
NO
.50
97
7
Frankfort plane
alignment tool