Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Planmeca Proline XC X-ray unit E-5
PANORAMIC X-RAY BEAM ADJUSTMENT
Technical Manual
Unscrew the two attachment screws of the quick connector mechanism cover with the 4mm
allen screw and slide the cover from its position.
NOTE
In case the X-ray unit is a pan/ceph unit with two fixed sensor heads, the cover is
attached to two attachment pins. Unscrew the two attachment screws with the 2.5mm
allen key and remove the cover by sliding it horizontally from its position.
Figure 9
Loosen the four screws located on the lock disc.
Figure 10
Cover
Lock disc