Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Planmeca Proline XC X-ray unit E-29
CALIBRATING PANORAMIC SENSOR HEAD
Technical Manual
On the
Segmentation
display
SWITCH OFF
the radiation by
deselecting
all the sectors with
the arrow fields. The sectors will turn dark.
Figure 56
The pmpcal2 window with text Waiting appears.
Figure 57
Touch the Ready field to drive the X-ray unit into the exposure position.
The text Exposure appears.
Figure 58
Take an exposure. After the exposure the message window shown below appears.
Figure 59
SWITCH ON
the radiation by touching the arrows of the deselected sectors again.