Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
E-74 Planmeca Proline XC X-ray unit
CEPHALOSTAT ADJUSTMENT
Technical Manual
Adjust the cephalostat head support position to horizontal position, if needed (see Fig. 138, 4
and Fig. 139 below)
Figure 139
Fasten the loosened screws and replace the covers.
NOTE
CHECK THE RADIATION BEAM ALIGNMENT ACCORDING TO THE INSTRUCTIONS
GIVEN IN SECTION 7.3 “Checking the cephalometric beam alignment” on page E-65
AFTER ADJUSTING THE HEAD SUPPORT POSITION. ADJUST THE BEAM, IF NEEDED.
REPEAT THE BEAM AND HEAD SUPPORT ADJUSTMENTS UNTIL THE RADIATION
BEAM POSITION AND HEAD SUPPORT POSITION ARE ACCEPTABLE.