Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Planmeca Proline XC X-ray unit E-39
PATIENT POSITIONING MECHANISM
Technical Manual
If the distance from the middle ball to the center of the tenth ball on the left-hand side is more
than 122mm and on the right-hand side less than 118mm then the patient positioning mecha-
nism must be moved away from the column (Fig. 80). Note that the distance between the line
on the underside of the alignment ruler and the y-line on the ball phantom must be less than
±1mm, see section “Adjustment in ball phantom’s y-direction” on page E-41.
Figure 80
If the distance from the middle ball to the center of the tenth ball on the left-hand side is less
than 118mm and on the right-hand side more than 122mm then the patient positioning mech-
anism must be moved towards the column (Fig. 81). Note that the distance between the line
on the underside of the alignment ruler and the y-line on the ball phantom must be less than
±1mm, see section “Adjustment in ball phantom’s y-direction” on page E-41.
Figure 81
Right-hand side
ST
AR
T
EN
D
PL
AN
ME
CA
No
.5
09
71
X-
LIN
E
Y-
LIN
E
P
atleft.eps
More than 122 mm Less than 118 mm
Right-hand side
STA
RT
END
PLA
NM
ECA
No.50971
X-LIN
E
Y-LINE
P
a
tr
ight.eps
Less than 118 mm More than 122 mm
Summary of Contents for Proline XC
Page 1: ...technical manual 10010942_10 EN...
Page 2: ......
Page 8: ...TABLE OF CONTENTS 6 Planmeca Proline XC with Dimax3 Technical manual...
Page 253: ...Planmeca Proline XC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Page 254: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline XC X ray unit Technical Manual...
Page 256: ......
Page 257: ......