Planmeca Proline XC X-ray unit F-1
Chapter
Technical Manual
F
PREVENTIVE MAINTENANCE
1 SYSTEM MAINTENANCE
1.1 Cleaning
CAUTION
When disinfecting the unit surfaces, always disconnect the unit from the mains. The
unit must not be exposed to gaseous disinfectants or explosive anesthetics. Never
spill any liquids into the unit.
The chin support and the Ceph cassette holder can be cleaned with NON-ALCOHOLIC disin-
fection solutions.
The bite piece of the chin rest can be removed and autoclaved. It can also be cleaned with
alcohol-based solutions. The patient support handles can be cleaned with alcohol-based
solutions.
Other unit surfaces can be cleaned with a soft cloth damped in a mild cleaning solution.
1.2 Operator’s checks
Exposure indicators
Confirm that the exposure indicator appears on the Graphic user interface and that the expo-
sure indicator light turn on in the remote exposure switch for the length of the exposure.
Exposure warning signal
Confirm that the buzzer comes on for the length of the exposure. If needed, adjust the tone of
the signal, see section “Sound tone” on page B-5.
Exposure switch
Confirm that the exposure switch requires continuous activation to maintain the exposure.
Releasing the exposure switch during the radiation should produce the error message on the
display. Make a visual check that the exposure switch spiral cable is not damaged.
Labels
Check that no labels are detached or worn and that they are all legible.
Summary of Contents for Proline XC
Page 1: ...technical manual 10010942_10 EN...
Page 2: ......
Page 8: ...TABLE OF CONTENTS 6 Planmeca Proline XC with Dimax3 Technical manual...
Page 253: ...Planmeca Proline XC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Page 254: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline XC X ray unit Technical Manual...
Page 256: ......
Page 257: ......