Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Planmeca Proline XC X-ray unit E-75
CEPHALOSTAT ADJUSTMENT
Technical Manual
7.7
Checking the second primary collimator position
You must now check the beam alignment with the second primary collimator in its position.
Attach the second primary collimator to its position.
Remove the sensor head and attach the sensor alignment tool to the cephalostat head sup-
port.
Enter the service settings mode. By touching the arrow down field below the list you can see
more features. Select the primary collimator mode from the Service settings list.
Figure 140
The display shown below appears. The number above the arrow fields indicates the collima-
tor number and the collimator is changed with the Next field. Select the collimator d.
Figure 141