Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Planmeca Proline XC X-ray unit E-45
PATIENT POSITIONING MECHANISM
Technical Manual
NOTE
Adjust the position of the rotating unit by holding the sensor head. Never use the tube
head to adjust the position.
Leave the ball phantom tool in position and place alignment tool adapter to the sensor head.
Place the alignment ruler in position between the primary collimator and the sensor alignment
tool as described earlier.
Figure 92
The line on the alignment ruler must line up with the x-line on the ball phantom (Fig. 93).
Figure 93
0° position
Enter the service settings mode.
STAR
T
END
PLANME
CA
No.50971
X-LIN
E
Y-
LIN
E
Dimax2_adj2.eps
START
END
PLANMECA
No.50971
X-LINE
Y-LINE
X-LINECA.EPS
Alignment ruler
Ball phantom
Tube head side
Sensor head side