Planmeca Intra X-ray unit 43
Technical Manual
PARTS REPLACEMENT & REPAIR
12. Coat the new arm cable with silicone, grease or similar. Pull from the tube head end of the
arm cable and simultaneously move the arm up and down.
Figure 20
NOTE
Make sure that the arm cable and the Dixi interconnection cable are not twisted when
you are pulling them into the arm.
13. Stop pulling when the new cable is 390-400 mm out from the opening of the tube head
support (figure 21, 1). Route the cable through the opening on the support frame (figure
21, 2).
Figure 21
Intr
a_PSU_D_08_11.eps
New arm cable
Old Dixi interconnection cable
390-400 mm
1
2
Summary of Contents for intra