Planmeca Intra X-ray unit 47
Technical Manual
PARTS REPLACEMENT & REPAIR
21. Connect the arm cable to the tube head PCB and connect the grounding lead to the
grounding point near the support frame opening.
Figure 28
22. Check that the arm cable remains in the groove of the tube head support when moving the
arm from one extreme position to another. Secure the arm cable to the support frame with
a cable tie.
23. Pull the cables through the extension arm and connect the cables to the generator
assembly connectors.
24. Replace the removed covers and plugs.
Feedback cable
Arm cable
Grounding lead of arm cable
Tube head grounding lead