PARTS REPLACEMENT & REPAIR
38 Planmeca Intra X-ray unit
Technical Manual
5. Remove the eight M4 DIN 934 attachment nuts and ø4.3 DIN 6798 washers from the
Generator PCB (figure 15, 1). Unscrew the two M4x8 DIN 84 screws (figure 15, 2). Lift the
generator assembly away from the wall adapter and open the two M3x6 DIN 912 screws
located behind the generator assembly frame (figure 15, 3).
6. Remove the Generator PCB.
Figure 15
7. Install the new Generator PCB in reverse order. Note, that the new Generator PCB is not
attached to the frame with the two M3x6 DIN 912 screws located behind the generator
assembly frame.
P6
P5
P7
P9
P3
N
N
L
L
Intr
a_PSU_D_08_9.eps
1
1
2
3
M4 DIN 934 nut
and M4 washer