Planmeca Intra X-ray unit 7
Technical Manual
GENERAL & TECHNICAL DATA
1.5
Stray radiation measurements for Planmeca Intra X-ray unit
The measurements are given for loading factors which result in the maximum local dose
per current time product. The loading factors include the highest selectable Xray tube
voltage.
PMMA phantom with a diameter of 160 mm and a height of 160 mm has been used for the
measurements. The phantom is compatible with the specification in international standard
IEC 60601-1-3.
The values were measured in the horizontal plane which was at the height of the centre of
the phantom and the X-ray source. The unit of measurement was air kerma per mAs
applied to the X-ray tube during normal use.
Exposure values:
70 kV / 8 mA / 0.5 s
Dose values are presented in nGy/mAs.
Figure 4