GENERAL & TECHNICAL DATA
2 Planmeca Intra X-ray unit
Technical Manual
1.3
Technical specifications
Generator
Constant potential, microprocessor controlled,
operating frequency 66 kHz
X-ray tube
Toshiba D-0711SB
Focal spot size
0.7 mm according to IEC 60336
Cone diameter
60 mm (2.36 in.)
Rectangular 33 x 43 mm (1.30 x 1.69 in.)
Max. symmetrical radiation field
ø 60 mm at SSD 200 mm
ø 60 mm at SSD 300 mm
according to IEC 806
Total filtration
min. 2 mm Al equivalent at 70 kV
according to IEC 60522
Inherent filtration
1 mm Al equivalent at 70 kV
according to IEC 60522
Anode voltage
50, 52, 55, 57, 60, 63, 66, 70 kV, ±2 kV
Anode current
8, 7, 6, 5, 4, 3, 2 mA ±(5% + 0.2 mA)
Target material
Tungsten
Target angle
16°
Exposure times
0.01-3.2 sec. ±(5% + 0.001 sec.), 26 steps
Reference current time product
8 mAs at 70 kV, 8 mA, 1 sec.
Lowest current time product
0.02 mAs at 2 mA, 0.01 sec.
Max. nominal anode voltage
70 kV
Power input
1000 VA
Max. electrical output
560 W at 70 kV, 8 mA
Electrical output at 0.1 sec.
560 W at 70 kV, 8 mA
Max. loading energy
1800 mAs/h at 70 kV
SID (SID = source - image
receptor distance)
min. 200 mm (8 in.)
SSD (Source-Skin Distance)
Standard/Long
Long with rectangular collimator
200 mm (8 in.)/300 mm (12 in.)
306 mm (12.04 in.)
Mains voltage
100 V~/110-115 V~/220-240 V~
Apparent resistance
0.3 ohms 100-115 V~ /
0.8 ohms 220-240 V~
Mains frequency
50/60 Hz
Fusing
units with 100V~ or 110-115V~ voltage
setting: 15AT, 250V, slow blow (6.3x32mm)
(special fuse, manufacturer Bussmann,
type MDA)
units with 220-240V~ voltage setting:
8AT, 250V, slow blow (6.3x32mm)
(special fuse, manufacturer Bussmann, type MDA)
Duty cycle
1:15, automatic control