GENERAL & TECHNICAL DATA
8 Planmeca Intra X-ray unit
Technical Manual
1.6
EMC information
WARNING
Use of any accessories and cables other than those specified in
Planmeca Intra X-ray unit’s documentation, with exception of
cables sold by Planmeca as replacement parts for internal
components, may result in increased emission or decreased
immunity of the X-ray unit.
WARNING
Planmeca Intra X-ray unit should not be used adjacent to or
stacked with other equipment. If adjacent or stacked use is
necessary, the Planmeca Intra X-ray unit should be observed to
verify normal operation in configuration which it will be used.
Guidance and manufacturer’s declaration - electromagnetic emissions
Planmeca Intra X-ray unit is intended for use in the electromagnetic environment specified
below. The customer or the user of the Planmeca Intra X-ray unit should assure that it is used in
such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
Planmeca Intra X-ray unit uses RF energy only
for its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
Planmeca Intra X-ray unit is suitable for use in
all establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies