RECALLING THE FACTORY PREPROGRAMMED
20 Planmeca Intra X-ray unit
Technical Manual
3
RECALLING THE FACTORY PREPROGRAMMED
EXPOSURE VALUES
The factory preprogrammed exposure values are in the Planmeca Intra user’s manual,
section 9 “EXPOSURE VALUES”.
Press and hold down any of the preprogrammed setting keys when switching the unit on.
The error code E.29 will appear on the time display.
Do not clear the error code by pressing the select key, but press the Occlusal exposure
key for 6 seconds. The factory preprogrammed values and density value 0 will be stored
into memory.
E.29