SERVICE MODE
16 Planmeca Intra X-ray unit
Technical Manual
2.8
Duration of the displays dimming time-out
Enter the service mode according to the instruction given in section 2.2 “How to enter/exit
the service mode” on page 12.
Press the parameter adjustment up key until the parameter number 17 appears on the kV
display.
The number indicating the time-out value (0, 1, 2, 3, 4, 5) is shown on the time display.
The time-out values are: 0 = no dimming, 1 = 1 min., 2 = 2 min., 3 = 5 min., 4 = 20 min. and
5 = 60 min. Factory default is 1.
Press the Select key until the number starts to blink, and the value can now be changed
with the parameter adjustment keys.
Accept the new setting by pressing the Select key.
Press the Mode key briefly to exit the service mode.
Parameter adjustment up -key
17
0
Parameter adjustment keys