Planmeca Intra X-ray unit 3
Technical Manual
GENERAL & TECHNICAL DATA
Mechanical data
Environmental requirements
External mains fuse recommendation
The recommendation for the external mains fuses are:
•
units with 100V~ or 115V~ voltage setting:
16A, time lag
•
units with 220-240V~ voltage setting:
10A, time lag
No other equipment should be connected to the same fused mains line as the x-ray unit. In
some countries an additional external fault current guard is also required.
Original manufacturer
PLANMECA Oy, Asentajankatu 6, FIN-00880, Helsinki, FINLAND
phone: +358-20-7795 500
1.4
User’s statement for Planmeca Intra
Radiation leakage technique factors
The maximum rated peak tube potential is 70 kV and the maximum rated continuous tube
current is 0.53 mA for the maximum rated peak tube potential.
Minimum filtration
The radiation port contains an added 1.0 mm aluminium filtration. The measured half-
value is 0.50 - 0.55 at 70 kV. The measured value corresponds to an aluminium equivalent
of 2.0 mm.
Rated line voltage
100, 110-117, 220-240 V~ ±10%. Line voltage regulation 10%.
Maximum line current
6.1 A at 230 V~, 12.2 A at 115 V~
Electrical classification
Class I
Type B
Weight
total 23 kg (51 lbs)
tube head 4.2 kg (9.3 lbs) with standard cone
4.5 kg (10 lbs) with long cone
Color
RAL 9016
Ambient temperature
ope5°C - +40°C
storage -10°C - +50°C
transport -10°C - +50°C
Humidity
25% - 75%
Atmospheric pressure range
700 hPa - 1060 hPa