Planmeca Intra X-ray unit 23
Technical Manual
PREVENTIVE MAINTENANCE
6. Move test leads to P1-connector pins 3 (negative mA) and 4 (positive mA).
Figure 6
7. Set multimeter to DC voltage mode with a measurement range of 100mV to 5V signall
evels. Take 2 seconds exposure with desired mA setting. Selected kV value has no effect,
however lowest possible kV should be used to minimise the amount of unnecessary
radiation. When the voltage reading has been stabilized, record it.
Actual tube current (in mA) = 2.5 * measured feedback voltage (in volts)
The resulting tube current should be within ±10% of the current indicated in the user
interface.