39
DRU10002 (REV 004 / JAN 2017)
WARNING:
Do not use an electrode shaft which appears to wobble, vibrate,
or tilt away from the axis of rotation while rotating. Such a shaft
is either improperly installed or physically damaged. Turn off the
rotator, disconnect electrical power, and remove the shaft
immediately.
AVERTISSEMENT:
N’utilisez pas un arbre d’électrode qui semble osciller, vibrer ou dévier de
l’axe de rotation pendant la rotation. Cet arbre est soit installé de manière
incorrecte soit endommagé physiquement. Éteignez le rotateur,
déconnectez
l’alimentation électrique et retirez l’arbre immédiatement.
If the shaft is rotating properly along the axis of rotation, then it is ready for use.
Some shafts are actually single-piece electrodes where the electrode tip is
permanently attached to the shaft. But most shafts are designed to accept a
variety of different tips. For these “shaft and tip” designs, the shaft may remain
mounted in the rotator, and changing the tip is a simple matter of unscrewing
one tip and then threading a new tip on to the shaft.
Figure 4.4: Installing a Tip on to a Shaft
Summary of Contents for MSR 636A
Page 2: ......