30
There are several holes in the floor of the
enclosure base
(4)
, which are threaded to
accept the side post
(6)
. Choose one of
these holes and install the side post in it. Then,
mount the laboratory clamp on to the side
post.
There are two short banana cables
(red and blue) which serve as jumpers
between the left and right brush
connections. Use the red cable to connect
the upper (red) pair of brush connections,
and use the blue cable to connect the lower
(blue) pair of brush connections, running the
wires behind the assembly as shown.
Insert one banana plug stud into the red
banana cable, and insert the other banana
plug stud into the blue banana cable. These
flat studs are an ideal place to make
connections using alligator clips.
The upper (red) jacks make electrical
contact with a rotating disk electrode (RDE)
or a rotating cylinder electrode (RCE) tip.
When using a rotating ring-disk electrode
(RRDE), the upper (red) jacks make contact
with the disk, and the lower (blue) jacks
make contact with the ring.
Remove the control unit
(10)
from the box
and place it next to the enclosure base. Plug
the male end of the motor control cable
(8)
into the motor cable connector on the back
of the control unit, and plug the female end
of the cable into the top of the motor unit.
Summary of Contents for MSR 636A