106
6.7
Changing the Input Rotation Rate Ratio
The rotation rate can be controlled by applying an external voltage signal to the
input jacks on the front panel of the control unit. The proportionality ratio used to
convert the applied voltage signal to the rotation rate can be set to one of three
different values: 1.0, 2.0 or 4.0 RPM/mV.
Normally, the value for the rotation rate ratio is selected to match the control
signal provided by a particular potentiostat. When shipped from the factory, the
MSR rotator is pre-configured with a ratio of 1.0 RPM/mV because this ratio is
compatible with Pine potentiostat systems.
WARNING:
Risk of electric shock.
Disconnect all power before servicing the rotator.
AVERTISSEMENT:
Risque de décharge électrique.
Déconnectez toutes les sources d’alimentation avant
de procéder à
l’entretien du rotateur
.
CAUTION:
Static electricity may damage electronic components.
Ensure proper grounding when handling static sensitive
components by wearing a grounding strap.
ATTENTION:
L’électricité statique est susceptible d’endommager les composants
électriques.
Veillez à disposer d’une mise à la terre appropriée lorsque vous manipulez
des composants sensibles en portant un ruban de mise à la terre.
With the power cord disconnected, remove the cover from the control unit.
Summary of Contents for MSR 636A
Page 2: ......