103
DRU10002 (REV 004 / JAN 2017)
Note:
A convenient certification sheet that can be used to record the
verification readings (from the previous step) can be found at
the end of this procedure.
Switch off power to the rotator and disconnect the power cord.
Use a small (5/64") hex key to loosen the hex screws in the motor coupling and
remove the shaft from the rotator.
Use the hex key to securely retighten the hex screws into the motor coupling.
Close the clamshell doors on the brush chamber and secure the latch.
Secure the enclosure around the rotator motor unit (see Figure 4.6).
WARNING:
Rotating shaft. Very high rotation rates!
The next part of the calibration procedure involves rotation rates
at or above 10000 RPM.
Before continuing, verify each of the following points:
•Verify that the shaft has been removed from the rotator
.
•Verify that the hex screws in the motor coupling are
tightened.
•Verify that the clamshell doors are closed and latched.
•Verify that the enclosure window is secured to all four pins as
shown below.
AVERTISSEMENT:
Arbre en rotation. Vitesses de rotation extrêmement élevées!
La prochaine partie de la procédure d’étalonnage implique des vitesses de
rotation supérieures ou égales à 10000 tr/min.
Avant de continuer, vérifiez chacun des points suivants :
•Vérifiez que l’arbre a été retiré du rotateur.
•Vérifiez que les vis hexagonales du raccord avec le moteur sont bien
serrées.
•Vérifiez que les portes de la coque sont fermées et verrouillées.
•Vérifiez que les quatre broches de fermeture de la fenêtre du boîtier sont
fermées tel qu’indiqué ci
-dessous.
Summary of Contents for MSR 636A