LPCR2 Service Process Work Instruction 

10 of 23 

 

Work

 

1

 

1

st

 

Verify

 

2

 

2

nd

 Verify 

 

Tool 

!

 

Caution 

 

Shock  
Hazard 

E

N

S

W

 

Orient 

 

Clean 

 

Scan 

 

Template: 7000743_B 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

2.  Handle or Carrying Case Replacement  

To replace the Handle or Carrying Case: 

2.1 

Remove three screws from the handle or carrying 
case. 

Note: For Handle, follow steps 2.2-2.3. For Carry Case, 
follow steps 2.4-2.9. 

Handle Installation 

2.2 

Place the device face down on the work surface 
and align Handle screw holes with back of 
device.  

2.3 

Install 3 screws (REF 21300-008145) to fasten 
Handle to device, using 7.8 in-lb.  

Carry Case Installation 

2.4 

Open front lid of Carry Case by pulling on the red 
fabric tab. 

2.5 

Place device into plastic frame of Carry Case, 
with device lid facing up. 

2.6 

Close front lid of Carry Case and flip Carry Case 
over. 

2.7 

Open back lid of Carry Case by pulling up on the 
black plastic tab. 

2.8 

Pre-thread the 3 screws one revolution into 
device by hand and then use torque driver to 
fasten the 3 screws on Carry Case to device, 
using the 7.8 in-lb p-set.  

2.9 

Close the back lid of Carry Case.

 

 

 

Handle (21300-008142) 

TO

R

Q

U

E

TO

R

Q

U

E

TO

R

Q

U

E

7.8 In-Lb 

Carry Case (11260-000046) 

3 Screws  

(21300-008145) 

 

Summary of Contents for LPCR2

Page 1: ...PR2 to LPCR2 Service Process Work Instruction Revised document type from Product Drawing to Manufacturing Specification Update document template to 7000743 revision B changed format to match new template requirements Removed and replaced all Physio Part Numbers with Catalog Numbers throughout Added definitions for LIFENET and LDA Removed Table of Contents Removed sections that are duplicated with ...

Page 2: ...nts Document No Title 3321515 Instructions Operating LIFEPAK CR2 3322174 Guide Wireless Setup LIFELINK LIFEPAK CR2 3323342 Test Procedure Manual PIP LPCR2 Service Personnel Qualifications These instructions are intended for authorized Service Representatives and Service Providers who have been trained to the Service procedure and who are familiar with the LIFEPAK CR2 defibrillator Preface This Ser...

Page 3: ...port Representative a Physio Control support team member dedicated to supporting customers with a full AED support service contract CR2 LIFEPAK CR2 Defibrillator DTx DT Express Data Transfer Software FSR Field Service Representative services products in the Field LIFELINK LIFELINKcentral AED Program Manager a user interface on the LIFENET system designed for Workplace and Community Customers to vi...

Page 4: ...n the LIFEPAK CR2 Defibrillator Workplace and Community A Physio control customer segment that represents public access locations for Automated External Defibrillators such as schools government agencies airports health clubs and workplaces Tools Required substitution of equivalent or better equipment allowed Computer Win 7 equivalent or better Cable USB2 0 A Male to Micro B L 5 5 ft 21300 008143 ...

Page 5: ... DEVICE DO NOT OPEN THIS DEVICE DOING SO MAY RESULT IN INJURY OR DEATH General Requirements Unless specifically noted all removed fasteners are to be discarded and new fasteners used for reassembly All removed parts with adhesive backing are to be discarded and new parts used for reassembly Unless specifically noted all removed parts except fasteners and adhesive backed parts are to be kept with t...

Page 6: ...he LPCR2 Operating Instructions CAT 26500 003800 or 26500 003754 Storing Wi Fi Settings on LPCR2 Refer to the Guide Wireless Setup LIFELINK LIFEPAK CR2 3322174 for Wi Fi configuration setup instructions This procedure describes how to store Wi Fi settings in the LPCR2 Preventative Maintenance Periodic maintenance inspection and testing of the device helps prevent and detect possible electrical and...

Page 7: ...00002 KIT HANDLE REPLACEMENT LPCR2 11260 000047 KIT CARRY CASE LPCR2 21300 008152 COVER USB PORT LPCR2 21300 008143 CABLE USB2 0 A MALE TO MICRO B L 5 5FT 26500 003645 TAG LOCAL EMERGENCY CONTACT NUMBER 911 MULTI Service Replacement Procedures Index Section Number Procedure Title Section 1 Lid Replacement Section 2 Handle or Carrying Case Replacement Section 3 USB Cover Replacement Section 4 Batte...

Page 8: ...acement NOTE LPCR2 Lid Replacement Kit REF 11512 000001 is used for Lid Replacement procedure NOTE If your defibrillator has a carrying case open the back of the carrying case to access the battery compartment 1 1 Remove the battery from the device 1 2 Open the lid 1 3 Push in one end of the lid hinge while rotating the lid to release the lid from the case Then Push in Top Lid 3316743 ...

Page 9: ...t Clean Scan Template 7000743_B 1 Lid Replacement Procedure Continued 1 4 Insert one end of the new lid into the small hole push in the other end of the lid and snap into place 1 5 Close the lid 1 6 Insert the battery 1 7 Open the lid voice prompts will begin 1 8 Close the lid to verify the readiness light is flashing in green every six seconds ...

Page 10: ...of device 2 3 Install 3 screws REF 21300 008145 to fasten Handle to device using 7 8 in lb Carry Case Installation 2 4 Open front lid of Carry Case by pulling on the red fabric tab 2 5 Place device into plastic frame of Carry Case with device lid facing up 2 6 Close front lid of Carry Case and flip Carry Case over 2 7 Open back lid of Carry Case by pulling up on the black plastic tab 2 8 Pre threa...

Page 11: ... 2nd Verify Tool Caution Shock Hazard E N S W Orient Clean Scan Template 7000743_B 3 USB Cover Replacement 3 1 Remove the USB plug from lower enclosure 3 2 Install the USB Cover onto Rear Case by inserting prong into small hole and push all the way in USB Plug 21300 008152 USB Plug 21300 008152 ...

Page 12: ...ntil 3 audible tones sound The defibrillator will automatically begin a self test NOTE The self test may take up to 2 minutes Do not open the lid until you hear the 3 tones Doing so will prevent the self test from completing NOTE If the same battery has previously been inserted the 3 tones will not sound 4 4 Verify that the green Readiness indicator is flashing every 6 seconds on the top of the li...

Page 13: ...1 Remove Electrode Tray by pressing lower tabs then pulling tray up and forward DO NOT USE HANDLE TO REMOVE ELECTRODE TRAY 5 2 Install electrode REF 11103 000026 by first sliding in the top edge then snapping in the two retaining clips in the bottom corners Note When the electrodes are used or expired the Readiness indicator will not flash until the electrode tray is replaced When a new electrode ...

Page 14: ...common failure modes 1 Readiness Indicator 1 1 If the readiness indicator is not flashing the possible causes will be the following Low battery power or expired electrodes Depleted battery Electrode tray not connected Operating temperature is too low or too high Defibrillator self test failed Follow the flowchart for the recommended corrective actions 1 2 If flashing but email notification indicat...

Page 15: ... a USB cable is required to connect the LPCR2 to the computer If the software cannot establish the connection between computer to the device the possible causes are Incorrect cable type use the USB cable shipped with the device REF 21300 008143 USB cable not properly connected Ensure the USB cable is firmly inserted into the USB port on the device COM port not assigned on Computer following direct...

Page 16: ...y mean that the signal strength at your location is not strong enough to connect the AED to the cellular network If possible move the AED installation and try again If you need assistance contact Physio Control Customer Support 3 1 Use a smart phone to test the Wi Fi signal strength and connection 3 2 If Wi Fi connection from the smartphone is working check that Wi Fi settings on the CR2 are corre...

Page 17: ...he voice prompts start and within 20 seconds after the voice prompts start If Device Ready or Device Not Ready is not heard that means the time window to accept the two button press has been missed Close the lid and try again There are additional voice prompts that are used during a cardiac arrest event See LIFEPAK CR2 Defibrillator Operating Instructions CAT 26500 003800 or 26500 003754 for the c...

Page 18: ...art LIFENET Check In after the completion of Connection Established This May Take Several Minutes Informs user to allow several minutes for updates to complete After the completion of Connection Established and before performing LIFENET Check In During LIFENET Check In device will beep every 5 seconds Update in Progress A software update is being downloaded and installed to the device Rebooting In...

Page 19: ...or support Connection Established The device has established a connection with LIFELINKcentral AED Program Manager or LIFENET System First HTTP response is an OK Start LIFENET Check In after the completion of Connection Established This May Take Several Minutes Informs user to allow several minutes for updates to complete After the completion of Connection Established and before performing LIFENET...

Page 20: ...OKLOGGEREVENT_HighChrgRate 1 0x40013 AOKLOGGEREVENT_LowChrgRate 1 0x40014 AOKLOGGEREVENT_OverVolt 1 0x40017 AOKLOGGEREVENT_MajorDiff 1 0x40020 AOKLOGGEREVENT_AokReset 1 0x40021 AOKLOGGEREVENT_SysReset 1 0x40040 AOKLOGGEREVENT_AdcCalibrationFail 1 0x40041 AOKLOGGEREVENT_WakeUpAlarm7DayTimeout 1 0x41700 SHOCKEVENT_OverVolt 1 0x41701 SHOCKEVENT_DumpTmrExp 1 0x41705 SHOCKEVENT_ChargeCmpltVcapChkFail 1...

Page 21: ...neFail 1 0x42512 STEVENT_PreampEcgRiseTmFail 1 0x42514 STEVENT_PreampEcgStepVoltFail 1 0x42516 STEVENT_PreampImpedInPhaseDcFail 1 0x42518 STEVENT_PreampImpedQuadratureDcFail 1 0x4251A STEVENT_PreampImpedBaseln1Fail 1 0x4251C STEVENT_PreampImpedPeakTm1Fail 1 0x4251E STEVENT_PreampImpedPeakVt1Fail 1 0x42520 STEVENT_PreampImpedDecyTm1Fail 1 0x42522 STEVENT_PreampImpedBaseln2Fail 1 0x42524 STEVENT_Pre...

Page 22: ... STEVENT_CellIntfCommFail 1 0x42558 STEVENT_PwrSuppliesAnalogFail 1 0x4255A STEVENT_PwrSuppliesGatedriveFail 1 0x4255C STEVENT_PwrSuppliesVddioFail 1 0x4255E STEVENT_PwrSuppliesVddaFail 1 0x42560 STEVENT_PwrSuppliesVdddFail 1 0x42562 STEVENT_PwrSuppliesSysFail 1 0x42564 STEVENT_LedsShockFail 1 0x42566 STEVENT_LedsPedFail 1 0x42568 STEVENT_RamConnectnFail 1 0x4256A STEVENT_WatchdogPetFail 1 0x4256C...

Page 23: ...Doc ID 3323299 Rev C Title LPCR2 Service Process Work Instruction 23 of 23 Work 1 1st Verify 2 2nd Verify Tool Caution Shock Hazard E N S W Orient Clean Scan Template 7000743_B System Context Diagram ...

Reviews: