Title
LPCR2 Service Process Work Instruction
Page
5 of 23
Doc Type
Service Process Document
Doc ID
3323299
Doc Rev
C
Doc State
Released
State Date
2/26/2020
Ref. Product DMR
:
Ref. PFMEA:
Ref. PFMEA Rev:
Template: 7000743_B
Warnings and Cautions:
THIS DEVICE HAS NO INTERNAL SERVICEABLE PARTS AND SHOULD NOT BE OPENED FOR ANY REASON.
LETHAL VOLTAGES MAY BE PRESENT INSIDE THIS DEVICE. DO NOT OPEN THIS DEVICE. DOING SO MAY RESULT IN
INJURY OR DEATH.
General Requirements:
•
Unless specifically noted, all removed fasteners are to be discarded and new fasteners used for reassembly.
•
All removed parts with adhesive backing are to be discarded and new parts used for reassembly.
•
Unless specifically noted, all removed parts, except fasteners and adhesive backed parts, are to be kept with the device being upgraded and
re-used to reassemble the device.
•
Inspect all removed components for damage during disassembly and replace all damaged parts with new parts.