4. Battery Replacement
NOTE
: If your defibrillator has a carrying case, open the
back of the carrying case to access the battery
compartment.
4.1
Remove the old battery.
4.2
Insert the new battery into the battery
compartment located on the back of the
defibrillator.
4.3
Wait until 3 audible tones sound. The defibrillator
will automatically begin a self-test.
NOTE
: The self-test may take up to 2 minutes.
Do not open the lid until you hear the 3 tones.
Doing so will prevent the self-test from
completing.
NOTE
: If the same battery has previously been
inserted, the 3 tones will not sound.
4.4
Verify that the green Readiness indicator is
flashing every 6 seconds on the top of the lid.
4.5
Open the lid and confirm that the voice prompts
are audible. Do not respond to the voice prompts
at this time.
4.6
Close the lid to turn off the device. Do not reopen
the lid unless necessary. Doing so will reduce
battery power.
Battery (11141-000163)