LPCR2 Service Process Work Instruction
6 of 23
Work
1
1
st
Verify
2
2
nd
Verify
Tool
!
Caution
Shock
Hazard
E
N
S
W
Orient
Clean
Scan
Template: 7000743_B
Service Information
The LPCR2 will not be opened for repair activities. Parts replacement is limited to the supply and replacement parts listed in Error! Reference source not found..
Before attempting to clean or service the device, the service technician should be familiar with the information provided in Error! Reference source not found..
A qualified service technician (see Error! Reference source not found.) should inspect any device that has been dropped, damaged, tampered, or abused to verify that
the device is operating within performance standards listed in the Performance Inspection Procedures (PIP), and that the leakage current values are acceptable.
Recycling Information
Recycle this product as described in Recycling Information in the LPCR2 Operating Instructions (CAT 26500-003800 or 26500-003754)
Storing Wi-Fi Settings on LPCR2
Refer to the Guide, Wireless Setup, LIFELINK, LIFEPAK CR2 (3322174) for Wi-Fi configuration setup instructions. This procedure describes how to store Wi-Fi settings
in the LPCR2.
Preventative Maintenance
Periodic maintenance, inspection, and testing of the device helps prevent and detect possible electrical and mechanical problems. Refer to the LPCR2 Operating
Instructions (CAT 26500-003800 or 26500-003754) for more preventative maintenance information such as battery maintenance and device self-test.
Performance Inspection Procedure (PIP)
Overview
: The Performance Inspection Procedure (PIP) is a set of manual and computer assisted test procedures used by authorized service representatives. The PIP
is used for an operational, closed-case evaluation of the LIFEPAK CR2 defibrillator to determine if the defibrillator is operating within the required specifications. The
PIP will be performed in a regularly scheduled preventative maintenance routine.
Note:
Always use the latest version of the PIP document, 3323342, to perform PIP.