42
BG 805 981 BE (0201) QMG422.oi
PE-CTRL
With
detect-TYPE:PENNING
OFF
ON
Penning switched off
Penning switched on (wait for ignition, pressure dependent)
PI-CH
With
detect-TYPE:PIRANI
0 / 1
Pirani channel to be measured
See
TEST-SERVICE
See
TEST-CS 422, -DSP
and -
QMS
QMA
Configuration input of the QMA type based on which the unit
recognizes the family
125
125
QMA with 6 mm rod system
400
400
QMA with 8 mm rod system
410
400
QMA with 16 mm rod system
430
400
QMA with 8 mm rod system (stainless steel)
See
TEST
The unit detects its modules automatically, as far as possible, and displays them.
See
TEST-CS 422, -DSP
and -
QMS
RANGE
With
amplif-MODE:FIX
and
detect-TYPE:FARAD
or
SEM
E-12 ... E-5
Manual electrometer range selection
RANGE-L
With
amplif-MODE:AUTO-D
and
detect-TYPE:FARAD
or
SEM
E-12 ... E-5
Lower search limit with
AUTO-D
See
INIT
400
Setting of the mass peak separation (resolution)
RESOL
With
detect-TYPE:FARAD, SEM, ION-CNT, EXTERN
OFF (0)
Integral mass spectrum (DC OFF)
1 ... 255
Mass peak separation. The peak width is approximately proportional to the set
number. Unit resolution at 20...30 (with QMH 400-1:
≈
100)
Often unit resolution suffices, that is, resolution of the adjacent peaks.
Decreasing the mass peak separation (larger number!) causes wider peaks and
higher sensitivity.
125
RESOL
Changeover Spectrum/Integral
OFF
Integral mass spectrum (DC OFF)
ON
Normal mass spectrum (DC ON)
Return from a submenu to the preceding menu
See
TEST-SERVICE
See
TEST-SERVICE
Channel related SEM high voltage
SEM
With
detect-TYPE:SEM, ION-CNT, EXTERN
SEM-HV (0)
The global value entered with
sem hv-SEM-VOLTAGE
is applicable.
1 ... 3500 V
Individual SEM high voltage for the selected measurement channel.
The individual SEM high voltage leads to long settling times and makes sense only
in special cases.
With High SEM (
config-SYSTEM-DETECT:H-SEM
) the minimum value is 750 V
(
→
18).
PE-CTRL (detect)
PI-CH (detect)
POS (config)
PRG-NR (config)
QMA (config-SYSTEM)
QMS (config)
QMS-HW (config)
RAM-T (config)
RANGE (amplif)
RANGE-L (amplif)
RESET (config)
RESOL (mass)
RETURN (....)
SELF (config)
SELF/CH (config)
SEM (detect)