BG 805 981 BE (0201) QMG422.oi
33
Parameters for operating the mass filter and for ion detection can be stored in up to
64 channels. When the measurement is performed the channels are processed
sequentially and cyclically.
channels
select
detect
mass
amplif
trip
output
aux
Control keys for all parameters of a
measurement channel
Choose a channel for parameter input.
All values entered under
channels
relate to this
selected
channel.
Determines the signal source (detector, e.g. Faraday or SEM).
Mass scan parameters such as mass number, speed, etc.
Measurement amplifier parameters, measurement ranges, Autorange/Fixrange....
Enable or skip the channel during the measurement operation and copy parameter
sets to different channels.
Parameters for analog output of measured values, e.g. linear or logarithmic.
Parameters of the switching functions of each measurement channel.
di/do
config
error
general
Keys for general settings such as configuration,
initialization, maintenance and service, and
processing of error messages.
Operation of DI-/DO-Bits.
Input and display of system and equipment configuration.
Error messages are displayed here in detail.
emiss
V1...V6
V7...
ion source
Parameters of the selected ion source set. The set
is chosen under
operation-ion src
and displayed
here.
400
Emission current and filament protection
5.2 Functions
Channels
group
select
detect
mass
amplif
aux
output
trip
General
group
di/do
config
error
ion source
group
emiss