BG 805 981 BE (0201) QMG422.oi
31
5 Description
In the basic version of the QMS 422 and on the QMI 422 the front panel contains
only a power switch and the corresponding LED.
The CS 422 operator console is available as an option. It can easily be retrofitted
and is highly recommended for learning purposes, e.g. for software development.
Even if you do not have a CS 422 you should carefully read the following
information. This will enhance your understanding also in computer mode.
With the
keys you can optimize the display contrast.
After a function has been selected, e.g. (
mass
key) the function (SETUP MASS),
its parameters (here
MODE
,
SPEED
...
RESOL
) and parameter values and the
measured value of the selected channel are displayed.
Parameter menus depend on the parameter state; in the picture below a
changeover to
MODE:SAMPLE
has been made
In computer mode the parameters are displayed for approx. 30 s when a key is
pressed. Input operations are disabled. Subsequently the measured value display
reappears automatically.
5.1 Operator console
Contrast setting
Parameter display
Up to 6 parameter names
and values
Channel selected for
input
Mass No.
Error messages and
warnings
Measured value numeric
...and as bargraph
Selected function