24
BG 805 981 BE (0201) QMG422.oi
In addition to 1. Configuration config-SYSTEM-DETECT:SEM
Module
Connector
Connection
l [m]
Module
Connector
Comments
HV420
HV–
High voltage
3
CP400
HV–
or 10 m
---
---
short SHV
---
CP400
HV+
short TNC
---
QMA
EP(FARAD)
if no EP
QC422
CP
control cable
3
CP400
QC
or 10 m
In addition to 1. Configuration config-SYSTEM-DETECT:H-SEM
Module
Connector
Connection
l[m]
Module
Connector
Comments
HV421
CD
remains open
HV–
High voltage
3
CP
HV–
or 10 m
HV+
High voltage
3
CP
HV+
or 10 m
short TNC
---
QMA
EP(FARAD)
If no EP
QC422
CP
control cable
3
CP400
QC
or 10 m
The
polarity
switch on the QME 125 must be in the "+" position.
Configuration config-SYSTEM-DETECT:FARAD
Module
Connector
Connection
l [m]
Module
Connector
Comments
QC422
QME
control cable QME
3
QME
QMS
or 10/20 m
QME
QMA
Analyzer
0.2
QMA
QME
ev. 6 m
EP422
Input
Meas. signal
---
QMA
EP
ev. 6 m
---
control cable
0.8
QME
EP
In addition to 1. Configuration config-SYSTEM-DETECT:CH-TRON
Module
Connector
Connection
l [m]
Module
Connector
Comments
QME
HV -
High voltage
0.3
QMA
HV -
ev. 6 m
In addition to 1. Configuration config-SYSTEM-DETECT:SEM
Module
Connector
Connection
l [m]
Module
Connector
Comments
QME
HV -
High voltage
0.3
QMA
HV -
ev. 6 m
EP422
Input
Meas. signal
short TNC
---
QMA
EP(SEM)
EP
in SEM mode
EP422
Input
Meas. signal
short TNC
---
QMA
EP
EP(SEM)
in Faraday
mode
---
control cable
0.8
QME
EP
4. SEM 217, HV 420
and ion counter
5. SEM 218, HV 421
and ion counter
4.6 Cabling with QMA 125
1. Faraday cup
2. Channeltron
3. 90° SEM