BG 805 981 BE (0201) QMG422.oi
19
Slots
1
Number per unit
max. 2
Supply
5 V / 0.45 A
Number of inputs
32
Input signals
24 VDC ±25% / 10 mA, low true
Switching threshold
10.6 ... 16 V
typ
Insulation
30 V
eff
/ 60 VDC to GND and between input groups
Protection
+35 V / -30 V continuous; 100 V max. 1 s
Connector
2, 16 inputs each, 32-pin DIN 41612 type C/2
Weight
0.24 kg
a c
1
16
J1/J2 pin assign-
ment viewed from
outside
n.c.
SOURCE14
SOURCE0
1
2
3
10
9
7
6
5
11
13
9
5
13
SOURCE4
SOURCE8
SOURCE12
IN 0
IN 15
IN 4
IN 8
IN 12
3
1
7
5
11
14
9
10
13
6
2
P
in
Example: Input N3 controlled with contact K
820
820
6.8V
Opto-
coupler
K
+
-
24V
a4
c4
DI 420
IN3
SOURCE3
SOURCE0..3 / 4...7 / 8...11 / 12...15 are internally connected, common plus poles
of the 4 insulated input groups. SOURCE15 has no pin of its own.
Slots
1
Number per unit
max. 3
Supply
5V / 0.4 A
Outputs
32, open collector Darlington drivers
Output signal
max. 28 VDC / 100 mA, U
sat
<1.7 V @ 100 mA, low true
Protection
Zener diode 33V/1W and parallel diode 1A
Insulation
30 V
eff
/ 60 VDC to GND and between groups
Connector
2, 16 outputs each, 32-pin DIN 41612 type C/2
Weight
0.24 kg
a c
1
16
J1/J2 pin assignment
viewed from outside
n.c.
RET14
RET0
1
2
3
RET10
9
7
6
RET5
11
13
9
5
13
4
8
12
OUT0
OUT15
OUT4
OUT8
OUT12
3
1
5
11
14
10
13
6
2
P
in
9
7
Example: Relay control with output OUT3
K
+
-
24V
a4
c4
DO 420A
OUT3
RET3
33
V
RET0..4 / 5...9 / 10...15 are internally connected, common negative poles of the 3
insulated outputs groups. RET15 has no pin of its own.
3.11 DI 420
Digital Input
3.12 DO 420A
Digital Output