56
BG 805 981 BE (0201) QMG422.oi
detect-TYPE:FARAD, SEM
output-
Output format
mass-MODE
AO-MODE,
MONITOR
amplif-
MODE
Decades
V/Decade
Formulas
SCAN-N
SCAN-F
LIN
FIX,AUTO
AUTO-D
1
10
a)
STAIR
PEAK-L
LOG 3D
FIX,AUTO
1)
AUTO-D
1)
3
3.333
b), c), d)
PEAK-F
ADJ-COARSE
3)
LOG 8D
AUTO
2)
AUTO-D
2)
8
1.25
e)
SAMPLE
LIN
FIX,AUTO
AUTO-D
1
10
a)
ADJ-FINE
3)
LOG 3D
FIX
3
3.333
b), c), d)
LOG 8D
AUTO
AUTO-D
8
1.25
e)
1)
only for SCAN-
SPEED
10...50 ms/u, with
STAIR
2... 5 ms/u
2)
only for SCAN-
SPEED
≥
100ms/u, with
STAIR
≥
10 ms/u
3) only at
mon
connector
LIN
a)
for Uo: 0...10 V:
I = 0.1 ×
RANGE
× Uo
0
1
2
3
4
5
6
7
8
9
1
10
X
0
1
2
3
4
5
6
7
8
9
10
-1
-1
10
X
=
RANGE
10
X-1
10
X-1
2 4 6 8
Formula a)
[A]
[V]
0,5
LOG 3D
b)
for Uo: 1V ... 10V
I = 10
-3
×
RANGE
× 10
0.3 Uo
c)
0.25V ... 1V
I = 1.333 ×10
-3
×
RANGE
× (Uo+0.5)
d)
0V ... 0.25V
I = 4 ×10
-3
×
RANGE
× Uo
Applicable with
amplif-MODE:AUTO:
RANGE = O-RNG
.
0
1
0
1
2
3
4
5
6
7
8
9
10
-1
10
X-3
2
4
6 8
Formula b)
[A]
[V]
0,5
10
X-2
1
0,25
1
2
4
6 8
10
X-1
1
2
4
6 8
10
X
6 8
Formula c)
Formula d)
10
X
=
RANGE
LOG 8D
e)
for Uo: 0...10 V:
I = 10
(0.8 Uo - 13)
0
1
2
3
4
5
6
7
8
9
10
-1
±10
-13
Formula e)
[A]
[V]
0,5
10
-12
10
-11
10
-10
10
-9
10
-8
10
-7
10
-6
10
-5
–10
-12
2 4 6 8
detect-TYPE:ION-CNT
output-
Output format
mass-MODE
AO-MODE,
MONITOR
output-
LOG-DEC
Decades
V/Decade
Formulas
SCAN, STAIR
LIN
---
1
10
g)
SAMPLE
LOG
3 DEC
3
3.333
h), i), k)
PEAK
ADJUST
3)
10 DEC
(>20 ms/u)
10
1
l), m)
Electrometer operation
Ion counting operation