BG 805 981 BE (0201) QMG422.oi
49
With
cycle-MODE
choose single channel
(MONO)
or multichannel
(MULTI)
mode
.
The measurement cycle is started and stopped with
run/halt
and its state is
indicated with the
mono, multi
and
halt
LEDs.
The number of measurement cycles is chosen with
cycle-CYCLES:
1...10000 or
repeating (0).
You can choose external control with
cycle-TRIG.
The measurement cycle is stopped. The unit measures in
mass-MODE:SAMPLE
(even if a different
MODE
has been entered) in the selected channel on the mass
defined with
MASS
(or
FIRST)
at the speed selected with
DWELL
(or
SPEED)
.
The switching functions are OFF, the signals on
elm, mon
and
AO
are available.
Single channel mode: enter
cycle-FUNCT:CYCLE
and
cycle-MODE:MONO
.
The unit measures in the measurement channel chosen with
select
. If the channel is
changed in the
run
state the measurement is cancelled and the new channel is started.
MONO
is suitable for measurement tasks in manual mode. A separate channel is
programmed for each task. This means that the parameters are continually avail-
able. You can quickly change the measurement task by changing the channel.
Multichannel mode: enter cycle-FUNCT:CYCLE and cycle-MODE:MULTI
The 64 channels can be programmed with any parameters. The channels located
between
cycle-BEGIN
and
cycle-END
are processed sequentially if they are not set
to
STATE:SKIP.
It is advantageous to first optimize each channel involved in MONO mode.
With the
aux-COPY
function load additional channels with the parameters of the
first optimized channel. Subsequently you only need to adapt a few parameters of
the individual channels.
To achieve the shortest measuring time the channels with identical detector type,
electrometer range and SEM voltage should directly follow each other.
The measurement cycle time CYCLE-TIME consists of the measurement and
pause times of the involved channels. It is measured by the built-in clock.
The clock starts with
run
, stops with
halt
and is displayed by pressing
cycle
. With
cycles-CYCLES:1
(no. of cycles =1) you measure the time for one cycle.
When the channel is changed in MULTI channel mode the data acquisition must
pause until the new measured value is stable. The pause time is set automatically
by the QC 422 based on
RANGE
,
FILTER
,
MASS
,
SEM
voltage and
detect-TYPE
.
The pause time is displayed under the softkey
amplif-PAUSE
. If you want to
achieve shorter cycle times or greater accuracy you can optimize it with
amplif-P-
CAL
. The minimum value is 1 ms with
P-CAL:0.0
.
You can reduce
P-CAL
in each channel until its measured value deviates
inadmissibly. The preceding channel should not have a measured value that is
almost identical, otherwise there is practically no transient response and the value
of
P-CAL
would be too small. Change e.g.
MASS
of the preceding channel by ½ u
to determine whether or not its measured value is without influence on the one of
the selected channel.
With the oscilloscope (triggering on the falling slope of
sync
) the transient
response during the pause can be observed at
elm
and (see also
55).
elm
sync
selected channel
6.8 Measurement cycle